Fractured Leader Italia implants: thoughts?
I am a Spanish Dentist. I placed Leader Italia, Tixos internal implants, a few years ago. Three implants on two of the patients have now fractured. In both cases, the patients had strong bite forces, although the protheses were different (a superior hybrid with two implants fractured and a second maxillar premolar with two crowns -without occlusion over the extension). Has anybody had a similar experience with the Leader Italia implants? What are your thoughts on this?
5 Comments on Fractured Leader Italia implants: thoughts?
New comments are currently closed for this post.
Vasileios Kouvas
4/22/2014
Dear colleague from Spain
I have the same bad experience with two TIxos internal hexagon implants from Leader Italia.The implant position was in 21,22 and 26 with a cement -retained metal -porcelain bridge.Three years later,my patient came to my office with complaints because of mobility of the restoration.The -x ray examination revealed that the broken implants were in the position of 22 and 26.The fracture line was at the same level of the apical level of the prosthetic screw.There was no evidence of strong chewing habits or bruxism.As you can supposed , the cost of the new implants and the prosthetic restoration was mine.. Until now I have the fear of the recurrence because I have finished a lot of cases with Tixos implants from Leader
Dennis Flanagan DDS MSc
4/29/2014
I have completed an in vitro research project on this issue. We found that implants can fracture at the cervical when loaded with more than 250 N in an off axial direction. This was also found in proprietary research by 3-I, poster presented at the AO meeting. It is a good idea to measure a pts bite force with pressure indicators from TekScan in South Boston, Mass. USA. This information is helpful in treatment planning.
Jaime
4/30/2014
1) Were machined components used ?
2) Don´t forget , even when the metal try-in in 2-dimensions looks great on radiograph , after firing porcelain , we have distortion . Southern Implants have passive abutments .
3) In case of high bite force keep the occlusal table narrower and decrease the inclined planes of the cusps .
José Vilches
5/5/2014
Hi. Abutments were machined from Leader. The fracture was in the middle of the implant, no in the platform, so I think would be a problem in the manufacurer.
José Vilches
10/29/2014
Hi. Another Tixos fractured. A #22 with a bridge screwed.